Calibration Key for Suretemp 690/692 Thermometers
Calibration Key for Suretemp 690/692 Thermometers
£72.79 ex.VAT
£72.79 ex.VAT
AVAILABLE TO BACKORDER - Buy now for despatch when stock arrives.
SKU: 06138-000 GTIN:00732094022858
More like this
More like this
Variants
Variants
Documents
Documents
Shipping
Shipping
FREE STANDARD SHIPPING available on all orders over £25 EX VAT.
Next working day - available until 8pm Monday to Friday
Weekend despatch - available until 12pm Sunday for MONDAY delivery (Next working day)
Sat and Sunday delivery service - available until 5pm Friday.
Why buy from Medisave?
Why buy from Medisave?
Medisave supplies health professionals, business and home users.
- Credit card, PayPal and Amazon Payments accepted
- Free delivery over £25 ex VAT
- NHS, school or company purchase order - email accounts@medisave.co.uk
- Laser engraving
- Delivery to private or commercial addresses
- Google Certified Shop
- Registered with the Medicines and Healthcare products Regulatory Agency (MHRA)
- ISO9001 quality management system + ISO 140001 environmental
- Established 1999
Need some advice? Speak to our friendly team on 0800 804 6447
Share
Calibration Key for Suretemp 690/692 Thermometers
Whenever the clinical accuracy of the thermometer is in question, verify the instrument calibration.
It is recommended the unit be tested annually.
This method uses a fixed value to verify that the device is calibrated properly.
1. Remove any probe currently connected to the SureTemp® Plus Thermometer.
2. Connect the SureTemp® Calibration Key.
3. Wait one minute for the device to enter Monitor mode and display the temperature.
4. The temperature should read 36.3 +/- 0.1ºC or 97.3 +/- 0.2ºF.
5. Remove the SureTemp® Calibration Key.
6. Connect the original probe back to the device.